User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Ballantine 3205B:Digital Micro-Ohmmeter DATE: 20-Aug-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 15 NUMBER OF LINES: 140 CONFIGURATION: Fluke 5700A STANDARD: EPC HS 5-500 STANDARD: SRI-SH 20 STANDARD: Leeds & Northrup 4222-B STANDARD: Leeds & Northrup 4221 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F 1.002 ASK+ X 1.003 HEAD INITIAL CONDITIONS 1.004 OPBR Are the Leeds & Northrup standard resistors available? 1.005 JMPT 1.009 1.006 MATH m[255] = -1 1.007 STD EPC HS 5-500 1.008 STD SRI-SH 20 1.009 STD Leeds & Northrup 4222-B 1.010 STD Leeds & Northrup 4221 1.011 DISP Set the UUT front panel controls as follows: 1.011 DISP RANGE switch ...................................... 20 1.011 DISP ohm/mohm switch ................................... mohm 1.011 DISP ZERO/OPERATE switch ............................... ZERO 1.011 DISP PULSE/CONT switch ................................. CONT 1.012 DRAW 8 78 2 78 100 5 66 55 1.013 DISP Adjust the ZERO adjust for a UUT reading of 0.000. 1.013 DISP [DRAW0,85,F] 1.014 DISP Release the ZERO/OPERATE switch to the OPERATE position. 1.015 HEAD {} 1.016 HEAD {* CALIBRATION ACCURACY TEST *} 1.017 JMP 2.001 1.018 EVAL 2.001 DISP Set the UUT to the 20mohm range. 2.002 DISP Connect the 0.01 ohm resistor to the UUT using the 2.002 DISP 4-wire method. 2.003 MEMI Use keyboard to enter calibrated value of Std. resistor: 2.004 MEM* 1000 2.005 MEME >1 2.006 MEMI Use keyboard to enter actual UUT reading in mohm. 2.007 MATH mem1 = m[255] 2.008 JMPT 2.011 2.009 ACC 2.5000mZ 0.01% 2.010 JMP 2.012 2.011 ACC 10.000mZ 0.01% 2.012 MEME 2.013 MEME <1 2.014 MEMC mZ 0.009U 3.001 DISP Disconnect the 0.01 ohm resistor from UUT. 3.002 DISP Set the UUT to the 200 mohm range. 3.003 DISP Connect the 0.1 ohm resistor to the UUT using the 3.003 DISP 4-wire method. 3.004 MEMI Use keyboard to enter calibrated value of Std. resistor: 3.005 MEM* 1000 3.006 MEME >2 3.007 MEMI Use keyboard to enter actual UUT reading in mohm. 3.008 ACC 100.00mZ 0.01% 3.009 MEME 3.010 MEME <2 3.011 MEMC mZ 0.09U 4.001 DISP Disconnect the 0.1 ohm resistor from UUT. 4.002 ASK+ R K 4.003 ASK- W 4.004 DISP Connect the 5700A and the UUT as follows: 4.004 DISP 5700A TO 3205B 4.004 DISP OUTPUT V ohm A HI ......... +V 4.004 DISP OUTPUT V ohm A LO ......... -V 4.004 DISP SENSE V ohm HI ........... +I 4.004 DISP SENSE V ohm HI ........... -I 4.005 5700 2000 1000.0mZ 0.9U 4W 5.001 5700 20 10.000Z 0.009U 4W 6.001 5700 200 100.00Z 0.09U 4W 7.001 5700 2000 1000.0Z 0.9U 4W 8.001 HEAD {} 8.002 HEAD {* PULSE FUNCTION TEST *} 8.003 JMP 9.001 8.004 EVAL 9.001 DISP Set the UUT to the 20 mohm range. 9.002 DISP Connect the 0.01 ohm resistor to the UUT using the 9.002 DISP 4-wire method. 9.003 DISP Depress the UUT PULSE/CONT switch. 9.004 DISP Momentarily depress the MEASURE switch. 9.005 MEMI Use keyboard to enter actual UUT reading in mohm. 9.006 MATH mem1 = m[255] 9.007 JMPT 9.010 9.008 ACC 2.5000mZ 0.01% 9.009 JMP 9.011 9.010 ACC 10.000mZ 0.01% 9.011 MEME 9.012 MEME <1 9.013 MEMC mZ 0.009U 10.001 DISP Disconnect the 0.01 ohm resistor from UUT. 10.002 DISP Set the UUT to the 200 mohm range. 10.003 DISP Connect the 0.1 ohm resistor to the UUT using the 10.003 DISP 4-wire method. 10.004 DISP Momentarily depress the MEASURE switch. 10.005 MEMI Use keyboard to enter actual UUT reading in mohm. 10.006 ACC 100.00mZ 0.01% 10.007 MEME 10.008 MEME <2 10.009 MEMC mZ 0.09U 11.001 DISP Disconnect the 0.1ohm resistor from UUT. 11.002 DISP Connect the 5700A and the UUT as follows: 11.002 DISP [27][91]1m 5700A TO 3205B 11.002 DISP OUTPUT V ohm A HI ......... +V 11.002 DISP OUTPUT V ohm A LO ......... -V 11.002 DISP SENSE V ohm HI ........... +I 11.002 DISP SENSE V ohm HI ........... -I 11.003 DISP Set the UUT to the 2000 mOhm range. 11.004 5700 1000.0mZ S 4W 11.005 DISP Momentarily depress the MEASURE switch. 11.006 MEMI Use keyboard to enter actual UUT reading in ohm. 11.007 MEME 11.008 MEMC mZ 0.9U 12.001 DISP Set the UUT to the 20 Ohm range. 12.002 5700 10.000Z S 4W 12.003 DISP Momentarily depress the MEASURE switch. 12.004 MEMI Use keyboard to enter actual UUT reading in ohm. 12.005 MEME 12.006 MEMC Z 0.009U 13.001 DISP Set the UUT to the 200 Ohm range. 13.002 5700 100.00Z S 4W 13.003 DISP Momentarily depress the MEASURE switch. 13.004 MEMI Use keyboard to enter actual UUT reading in ohm. 13.005 MEME 13.006 MEMC Z 0.09U 14.001 DISP Set the UUT to the 2000 Ohm range. 14.002 5700 1000.0Z S 4W 14.003 DISP Momentarily depress the MEASURE switch. 14.004 MEMI Use keyboard to enter actual UUT reading in ohm. 14.005 MEME 14.006 MEMC Z 0.9U 15.001 HEAD {} 15.002 END